HPHT Filter Press GGS42-2A

product_banner
HPHT  Filter Press GGS42-2A-2703
HPHT  Filter Press GGS42-2A-2700
HPHT  Filter Press GGS42-2A-2701
HPHT  Filter Press GGS42-2A-2702
HPHT  Filter Press GGS42-2A-2699
HPHT  Filter Press GGS42-2A-2698
GGS42-2A
_0003_IMG_20200811_155640
_0005_IMG_20200811_155623
_0006_IMG_20200811_155133
_0002_IMG_20200811_155825_1
_0001_IMG_20200811_155839

HPHT Filter Press GGS42-2A

Product Introduction

HPHT filter press named GGS42-2A is a professional instrument, used to measuring the filtration amount of drilling fluid and cement slurry under simulative deep well conditions (high temperature and high pressure), and also make filter cake after filtering under the same conditions. The filter is mainly composed of main body (base, stand bracket, heating system),tee valve, drilling fluid cup, high pressure filtrate receiver and QG-76 pressurized manifold, etc.

This filter has the characteristic of high temperature control precision, good repeatability, small error, simple operation, accurate measurement and etc. It has widely used for measuring the filtration amount of drilling fluid and cement slurry in oil fields, scientific research institutes, laboratories and other departments.


Working principle

Pour the quantitative drilling fluid samples into the drilling fluid cup. After sealing, put the drilling fluid cup into the heater and heat it. At the same time, inflated and pressurized the drilling fluid cup quantitatively through the pressurizing pipe. Under some kind of temperature and pressure conditions, pressurize the sample to produce the filtrate pass through the filter screen or filer paper in the drilling fluid cup, and then evaluate the filtration performance of drilling fluid by measuring the volume of filtrate.


Models and Specifications

Model

Name

Configuration

GGS42-2A

HPHT filter press

PID   temperature controller


The main technical parameters

 NO.

Item

Technical   specifications

1

Power supply

AC(220±11)V 50Hz

2

Rated power

400W

3

Maximum operate pressure

4.2MPa

4

Saver pressure

0.69MPa

5

Maximum   drilling fluid volume

175mL

6

Filter area

3.5 in2   22.6 cm2

7

Gas supply

Nitrogen, carbon dioxide(No impurities such as oil and water)

8

Operate   temperature

Normal to 150℃±3℃

9

Net weight/Gross weight

13.26kg/22.6kg

10

Dimensions

260×250×560mm

11

Packing dimensions (wooden box)

570×435×275mm    310×290×285mm


Inquiry now